Product Alert: t:slim X2 Insulin Pump - AMSL Diabetes

Product Alert: t:slim X2 Insulin Pump

What is the problem?

Certain versions of the t:slim X2 insulin pump have a wiring problem with the speaker. This may result in a failure which will appear on the pump screen as Malfunction Code 16-0x20E6, also known as “Malfunction 16”. When Malfunction 16 occurs, insulin delivery stops, and the pump becomes non-operational.

Image 1: t:slim X2 insulin pump

ARTG Number

Catalog number

Affected Serial Numbers

304681

1002717 

Multiple Serial Numbers: 5317 unique serial numbers identified in Australia

What are the risks?

 

The problem could:

  • •      Stop insulin delivery and disconnect the pump from the CGM and mobile app. 
  • •      Result in loss of real-time glucose data and trends.  
  • •      Lead to hyperglycaemia if not addressed promptly. 
  • •      Require patients to seek medical intervention or hospitalisation in serious cases.

What should I do?


If Malfunction 16 has not occurred:

•      Continue using your pump as per the user guide. 
•      Install the Tandem t:slim mobile app and enable push notifications. This ensures that if the problem does occur, you will be notified by the app.
•      If using a Dexcom sensor, use the Dexcom mobile app as a backup source for CGM alerts and alarms.

If a Malfunction 16 occurs:

1.    Acknowledge the alert by pressing “Silence Alarm.”
2.   Contact AMSL Technical Support at 1300 851 056, who will arrange for a replacement pump.
3.   In the meantime, switch to your backup insulin delivery method.

You should continue to monitor your blood glucose levels regularly and ensure a backup insulin method is available as per the pump training and user guide.

Further information


Please contact AMSL on 1300 851 056 for further information.

A software update is in development that will improve early detection of speaker failure and introduce persistent vibration alerts. Pump users will be notified when this software update becomes available.

Back to top